User`s guide
310 Agilent X-Series Signal Generators User’s Guide
Real–Time Phase Noise Impairments (Option 432) 
Real–Time Phase Noise Impairment 
Real–Time Phase Noise Impairment
This feature lets you degrade the phase noise performance of the signal generator by controlling two 
frequency points and an amplitude value. The signal generator adds this phase noise to the phase 
noise normally produced by the signal generator. This feature appears in each of the arb formats and 
as a stand–alone menu. While the following figure shows how to access the controls using both the 
stand–alone menu and the Dual ARB player, the location and softkeys within each arb format is the 
same as for the Dual ARB player.
Figure 12-1 Stand–Alone and Dual ARB Player Real–Time Phase Noise Softkeys
The end frequency for the flat 
mid–frequency characteristics.
The start frequency for the flat 
mid–frequency characteristics.
For details on each key, use key help 
as described on page 44. 
The display updates even 
when the feature is off.
For an explanation on these 
settings see
“The Agilent X-Series Phase 
Noise Shape and Additive Phase 
Noise Impairments” on page 311.
Flat Mid–Frequency Area
Denoted by f1 and f2
This softkey is 
available in all arb 
formats.
The amplitude for the flat 
mid–frequency characteristics. 
(Lmid stands for the level 
amplitude of the mid–frequency 
characteristic.)
NOTE:  The phase noise impairment is added to the base phase noise of    
 the signal generator.
Stand–alone menu
Use to add phase noise to the 
CW output without having to 
play a waveform.