User`s guide
 Agilent X-Series Signal Generators User’s Guide 351
 Custom Digital Modulation (Option 431)
Using the Arbitrary Waveform Generator
Creating a Multicarrier Digital Modulation Setup
1. Press  Preset.
2. Press  Mode > ARB Custom Modulation > Multicarrier Off On to On.
3. Press  Multicarrier Setup > Select Carrier and Initialize Table > Carrier Setup > EDGE > Done.
Modifying Carrier Frequency Offset
1. Highlight the Freq Offset value (500.000 kHz) for the carrier in row 2.
2. Press  –625 > kHz.
Modifying Carrier Power
1. Highlight the Power value (0.00 dB) for the carrier in row 2.
2. Press  –10 > dB.
You now have a custom 2–carrier EDGE waveform with a carrier at a frequency offset of 
−625 kHz and a power level of −10.00 dBm, as shown in the following figure.
Generating the Waveform
Press  Return > Digital Modulation Off On.
This generates a waveform with the custom, multicarrier, EDGE state created in the previous 
sections. The display changes to Dig Mod Setup: Multicarrier (Modified). During waveform 
generation, the DIGMOD and I/Q annunciators appear and the new custom, multicarrier, EDGE state 
is stored in volatile memory. The waveform is now modulating the RF carrier.
For instructions on storing this custom, multicarrier, EDGE state to non–volatile memory, see 
“Storing a Custom Multicarrier Digital Modulation State” on page 352. 
Configuring the RF Output
1. Set the RF output frequency to 890.01 MHz. 
2. Set the output amplitude to −10 dBm. 
3. Press  RF On/Off.
The custom multicarrier EDGE signal is now available at the RF OUTPUT connector.
If Digital Modulation is already on, you 
must press Apply Multicarrier to apply 
the changes and generate a new 
custom multicarrier digital modulation 
waveform based on the updated 
values.
For details on each key, use key 
help as described on page 44.