User`s guide
 Agilent X-Series Signal Generators User’s Guide 41
Setting Preferences & Enabling Options
Hardware Assembly Installation and Removal Softkeys
 Hardware Assembly Installation and Removal Softkeys
Utility > More 2 of 2 > Service
For details on each key, use key help 
as described on page 44.
Verify output attenuator 
operation using a power 
meter at the RF Output.
Select either Enhanced Factory 
Calibration or Factory Calibration to 
calibrate your instrument.
The Enhanced Factory Calibration 
requires a spectrum analyzer in 
addition to a power meter.
The Factory Calibration uses only a 
power meter.
Whether a softkey 
is available depends 
on the model of signal 
generator.
Whether a softkey 
is available depends 
on the model of signal 
generator.