User`s guide
 Agilent X-Series Signal Generators User’s Guide  51
 Basic Operation
Configuring a Swept Output
Figure 3-4 Sweep Softkeys
Sweep without waiting for  
a trigger at each point.  
Point Trigger pauses for the dwell 
time prior to the first sweep. See 
page 126.
Apply a TTL/CMOS signal to the 
Trigger In connector.
Periodically issue a trigger event to 
whatever selects it as a source.
Using timer trigger with single 
sweep results in a delay prior to the 
first sweep. See page 126.
Trigger on a remote command.
Down sweeps from 
stop to start 
frequency/amplitude.
Up sweeps from start 
to stop 
frequency/amplitude.
page 55
The selected sweep type determines the displayed parameter.  
Selecting step sweep also displays the step spacing (Lin or Log).
On vector models:
Available when 
Sweep Type = List
page 59
During a sweep, the swept parameter (frequency, amplitude, or both) turns grey and changes as the parameter sweeps.
Progress Bar: Note that very fast sweeps 
can appear to sweep randomly or 
backward.
page 52
For details on each key, use key help as described on page 44.
 page 52