 Agilent X-Series Signal Generators User’s Guide  59
 Basic Operation
Modulating the Carrier Signal
Example: Manual Control of Sweep
1. Set up either a step sweep (page 53) or a list sweep (page 56).
2. In the Sweep/List menu, select a parameter to sweep: Press Sweep > parameter > Return.
3. Select manual mode: Press More > Manual Mode Off On.
When you select manual mode, the current sweep point becomes the selected manual point.
4. If it is not already on, turn the RF output on: Press RF On/Off.
5. Select the desired point to output: Press Manual Point > number > Enter.
The progress bar changes to indicate the selected point.
6. Use the knob or arrow keys to move from point to point. As you select each point, the RF output 
changes to the settings in that selection.
Modulating the Carrier Signal
To modulate the carrier signal, you must have both 
• an active modulation format 
and
• modulation of the RF output enabled
Example
1. Preset the signal generator.
2. Turn on AM modulation: Press AM > AM Off On (requires Option UNT).
You can turn on the modulation format before or after setting signal parameters.
The modulation format generates, but does not yet modulate the carrier signal.
Once the signal generates, an annunciator showing the name of the format appears, indicating 
that a modulation format is active.
The parameters of the selected sweep point (point 3 
of 6 in this example) define the signal available at the 
RF Output connector.
When you enter a manual point, 
the progress bar moves to and 
stops at the selected point.
The SWMAN annunciator 
indicates that the sweep 
is in manual mode.
When you turn manual mode on, the 
current sweep point becomes the 
manual point.
The currently sweeping 
parameter stops and 
displays the selected point.