Troubleshooting the 
Agilent V24/V26 Series
 Green LEDs on the keypad (4 arrow and Confirm keys) are on momentarily when 
power is applied to the instrument. 
)LJXUH.H\SDG*UHHQ/('V%DWWHU\PRGHOVKRZQ
 There are 3 green LEDs on the power supply which may be observed by looking through 
the air vents on the left side of the instrument. With power applied, during normal system 
operation, all 3 LEDs should be on.
For the battery model, follow the right-hand column in the table. Otherwise follow the 
leviathan column.
7DEOH
0RGHOV99&9&
0RGHO9&79&7%DWWHU\YHUVLRQ
1RWHThe Model V24 (monochrome) uses power supply 
M1204-63000 (option #030). The Model V24C/V26C 
(color) uses power supply M1204-63000 (option #031). 
The option number is on the mainframe serial number 
label on the outside of the instrument.
1RWHThe Model V24CT/V26CT (battery) uses the 
power supply M1204-61086. The option number is on 
the mainframe serial number label on the outside of the 
instrument.
)LJXUH9LHZRI3RZHU6XSSO\*UHHQ/('V
0RGHO9DQG9&9&
)LJXUH9LHZRI3RZHU6XSSO\*UHHQ/('V
0RGHO9&79&7
Delayed
Realtime
Record
Confirm
Main
Screen
Red
Alarm
Yellow
Alarm
Silence/
Reset
Suspend
Record
ON
OFF 0
Monitor
Setup
Module
Setup
Other 
Patients
Procedures
1
Alarms
GREEN LED
GREEN LED
Trends/
Calcs
ARROW KEYS
CONFIRM
HARDKEY
Battery
Charging
Battery
Charged
AC
Power
 LEDs FOR
BATTERY MODEL
+5VDC
LED
+12V VCD
LED
+60 VDC
LED
18VDC 5VDC 12/60VDC