Specifications
5-6 Troubleshooting Plug-in Modules
$JLOHQW0$996HULHV6HUYLFH*XLGH
Troubleshooting Plug-
in Modules
❏  If tcpO
2
/tcpCO
2
 UHDGLQJVDUHTXHVWLRQDEOH, perform the following actions:
Perform a tcpO
2
 zero check and tcpCO
2
 gain check for the transducer. These checks tell you 
whether the tcpO
2
/tcpCO
2
 transducer is working correctly.
7HVW(TXLSPHQW
The following list gives you the test equipment recommended to perform the specification test on 
the tcpO
2
/tcpCO
2
 plug-in:
• 1 Calibration Unit (Agilent 15210B or Radiometer TCC3) 
• 1 Calibration gas bottle of Cal 1 gas (20.9% O
2
, 5% CO
2
) (see Chapter 11  “Replacement 
Parts”)
• 1 Calibration gas bottle of Cal 2 gas (0% O
2
, 10% CO
2
) (see Chapter 11  “Replacement 
Parts”)
• 1 Accessory Kit (15209-60010) 
• 1 tcpO
2
/CO
2
 transducer (Agilent M1918A or Radiometer transducer) 
• 1 Calibration Tubing (M2205A)
3URFHGXUH
6WHS Ensure correct function of the Calibration Unit.
6WHS Make sure that you are not in Service mode.
6WHS Set the transducer temperature to 43.0
o
 C. 
6WHS Using a calibration unit (Agilent M1918A or Radiometer calibration unit TCC3) and Cal 
1 gas, calibrate the module as described in the tcpO2/tcpCO2 Module sectionof the
8VHU¶V5HIHUHQFH0DQXDO
6WHS Check and note down the displayed values for tcpO
2
 and tcpCO
2
. The values should not 
fluctuate more than 2 mmHg (0.3kPa) in 30 seconds. 
6WHS If the readings fluctuate more than 2 mmHg (0.3kPa), re-membrane the transducer 
ensuring that the old electrolyte is thoroughly removed and check wether you need to 
activate the transducer. The tcpO2/tcpCO2 Module sectionof the 
8VHU¶V5HIHUHQFH0DQXDO
tells you how to do this. 
6WHS Repeat steps 4 and 5. If the test fails again, the transducer has unacceptable drift and has 
to be replaced. Repeat steps 4 and 5 with new transducer.
6WHS Replace the Cal 1 gas cylinder on the calibration unit with the Cal 2 gas cylinder. Turn 
on the gas supply of your calibration unit.










