CO
2
 OCCLUSION Sidestream flow between 50-80 
ml for more than 10 minutes or
Tubing locked in pump system, 
tubing or filter to patient.
Replace tubing.
CO
2
 SIDESTRM 
MALF
Sidestream malfunction in the 
transducer or module.
Replace module.
CO
2 
SIDESTRM OFF Pump automatically has turned 
off after 60 min of C02 < 3 
mmHg or apnea.
Restart Sidestream in CO
2
 setup.
7DEOH,1230HVVDJHIRU&2 0RGXOH