Specifications
5-26 Troubleshooting Plug-in Modules
$JLOHQW0$996HULHV6HUYLFH*XLGH
Troubleshooting Plug-
in Modules
7DEOH7DVN:LQGRZPHVVDJHVIRUWKH%ORRG$QDO\VLV0RGXOH
0HVVDJHLQ
7DVN
:LQGRZ
,123RU
SURPSW
PHVVDJH
(UURU
&RGH
3RVVLEOH&DXVH &RUUHFWLYH$FWLRQ
Temperature out 
of range
BLOOD-
ANL ROOM 
TEMP
2 The room temperature 
is too cold or warm - 
operating range is 16 to 
30°C (64 to 86°F).
The filter is blocked or 
the ventilator fan in the 
module malfunctions.
Ensure that ambient tem-
perature is within the 
operating limits and leave 
time for the module to 
acclimatize after trans-
port or storage.
Inspect the filter and 
clean it if necessary.
If the message still 
appears, call your service 
representative.
New Software 
Installed, Use 
Simulator 
3 The software or CLEW 
was recently updated.
Run the electronic simu-
lator.
Analysis Inter-
rupted, Use 
Other Cartridge
4, 8 Module has been 
pulled out of the rack 
during a measurement 
and then replaced.
Wait for the locked LED 
to go off then remove the 
cartridge and start the 
measurement again with 
a new cartridge.
Invalid Date, 
Check clock
BLOOD-
ANL CON-
FIG
11 Implausible date in 
V24/V26.
Check date settings in the 
V24/V26 and make sure 
they are correct.
Invalid or 
Expired CLEW, 
See Manual
BLOOD-
ANL CON-
FIG
12, 13 Invalid date code in 
V24/V26.
CLEW version maybe 
expired.
CLEW update has not 
been successfully fin-
ished.
Check date settings in the 
V24/V26 and make sure 
your date settings are cor-
rect.
Refer to the application 
note of the current 
CLEW version.
Reinstall the current 
CLEW version.
Invalid Customi-
zation Data
BLOOD-
ANL CON-
FIG
14 Customization update 
unsuccessful
Reinstall the current Cus-
tomization version.
Invalid Software 
Detected
BLOOD-
ANL CON-
FIG
53 Wrong software type 
installed.
Update System with cor-
rect type of software.