Specifications
$JLOHQW0$996HULHV6HUYLFH*XLGH
5HSDLULQJWKH$JLOHQW0RGHO9&79&7
Repairing the Agilent 
Model V24CT/V26CT
5HPRYLQJWKH3DUDPHWHU0RGXOH5DFN
 Press the power switch to OFF and disconnect the AC power cord. 
 Remove Lead Acid batteries. 
 Remove parameter modules from parameter module rack by pressing up on retaining tab 
and sliding parameter modules from rack.
 If you have a 6-slot rack docked to the V24CT/V26CT, press and hold retaining clip for-
ward. Use other hand to hold monitor steady. Pull the parameter module rack from the 
monitor.
Front Housing Removal Procedures
For the rest of the front housing removal procedures see Chapter 9.










