Specifications
$JLOHQW0$996HULHV6HUYLFH*XLGH
7HVWLQJ0DLQWDLQLQJWKH$JLOHQW996HULHV
Testing & Maintaining 
the Agilent V24/V26 
Series
Safety Tests
For Europe and Asia-Pacific:
according to IEC 601-1:1988+A1+1991+A2:1995 = EN60601-1:1990+A1:1991+A2:1995
For the US:
according to UL 2601-1
The test procedures outlined in this section are to be used only for verifying safe 
installation or service of the product in question. 
The set-ups used for these tests and the acceptable ranges of values are derived from 
local and international standards but may not be equivalent. 
These tests are not a substitute for local safety testing where it is required for an 
installation or a service event. 
If using the Metron Safety tester, use your local regulation to perform the test, for 
example, in Europe IEC601-1/IEC601-1-1, and in the U.S. UL2601-1. The Metron 
Report should print results with the names listed below, along with other data.
The Safety Tests required to ensure that the V24/V26 Series is in safe working order 
are detailed in the following tables.
7DEOH7HVWDQG,QVSHFWLRQ5HTXLUHPHQWV±6DIHW\
7HVW%ORFN
1DPH
7HVWRU³,QVSHFWLRQ´WR3HUIRUP ([SHFWHG7HVW5HVXOWV :KDWWR5HFRUGRQ
6HUYLFH5HFRUG
6DIHW\ System Safety Test Enclosure Leakage Current /
Normal Condition:
System Safety Test Enclosure Leakage Current / 
Single Fault Condition
Normal Condition 
maximum leakage 
current = x1. 
x1 <= 100 
µA
IEC601-1, UL2601-1 
(UL544 test not 
required)
Single Fault 
maximum leakage 
current = x2.
x2 is
<= 500 
µA (IEC601-1)
or
<= 300 
µA (UL2601-1)
or
<= 100 
µA (UL544)
63[[RU
6)[[
where P = Pass
and F = Fail
6DIHW\ Protective Earth With mains cable:
Maximum impedance = 
x <= 100 mOhms
All authorities 
(IEC601-1, UL2601-1, 
UL544)
63[RU6)[
where P = Pass
and F = Fail