Testing & Maintaining 
the Agilent V24/V26 
Series
*HQHUDO,QVSHFWLRQVRIWKH6\VWHP
,QVSHFWWKH6\VWHP+DUGZDUH
6WHS Examine the exterior of the unit for cleanliness and general physical condition. 
Ensure that plastic housings are intact, that all hardware is present and tight, and 
that there are no spilled liquids or other serious abuse.
6WHS Inspect connectors of the Rack for cleanliness and good condition. Ensure that the 
Modules are locked into place and do not slide out without first releasing the lock-
ing plate on the bottom of the Module.
6WHS Ensure all labeling is present and legible.
6WHS Inspect all accessories external to the system such as transducers, referring to the 
manufacturer’s documentation.
6WHS Verify the backlight tube brightness is adequate. Replace the backlight tube if nec-
essary. Agilent recommends replacement of the backlight tube every 18 months of 
continuous use. 
,QVSHFWWKH&DEOHVDQG&RUGV
6WHS Examine the line/power plug for damage. Ensure that the prongs of the plug do not 
move in the casing.
6WHS Inspect the line/power cord for signs of damage. If damaged, replace the entire 
cord. Ensure that any new line/power cord or plug is fitted with the same polarity as 
the old one.
6WHS Inspect the interconnecting cables between the module rack and the mainframe for 
general condition. Clean or replace contacts or housings are damaged, corroded or 
dirty. Examine them carefully to detect breaks in the insulation. Ensure that the 
cable connectors are properly engaged and that any connector screws are fully 
tightened.
6WHS Inspect the patient cables and leads and their strain reliefs for general condition. 
Examine cables carefully to detect breaks in the insulation and to ensure that they 
are gripped securely in the connectors of each end to prevent rotation or other 
strain. 
6WHS Flex the patient cable near each end to verify that there are no intermittent faults.