User manual
Identifying ECG Problems
7-3
Chapter 7
Troubleshooting
If the Recording Won’t Start 
If you press  and the recording doesn’t start, investigate the following 
possibilities: 
• Is the cardiograph turned on?
The screen should be on.
• Is the Power Indicator on? 
If the cardiograph is plugged into AC power via the AC/DC Adapter and the 
Power Indicator is not on, check the outlet.
• Is the battery adequately charged? 
The  symbol should not be displayed on the screen. 
• Is there an error message?
See “Error Messages” later in this chapter for more information.
If the cardiograph still won’t operate, perform the following steps:
1. Switch the cardiograph off with the   key. (If the cardiograph cannot be 
turned off, follow Appendix A to remove the battery, wait for about 20 seconds, 
and then install the battery. Connect the cardiograph to AC.)
2. Wait 20 seconds or more and then switch the cardiograph back to On. 
3. Press . If the cardiograph turns itself off, the battery is not operating prop-
erly. 
If the cardiograph still won’t operate, call your local Agilent service representative. 
Intermittent or Jittery 
Waveform
Poor electrode contact. Dry electrodes.
Faulty lead wires.
Clean the electrode sites. Reapply electrodes.
Replace faulty patient cable.
Table 7-2 ECG Problems and Solutions (Continued)
Problem Likely Cause Possible Solution