Identifying Transmission Problems
7-8
Transmission stopped 
unexpectedly. X of N ECGs sent.
Modem was disconnected. 
Press any key to continue.
Problem with telephone line Check that the modem is 
connected to the telephone line. 
Verify that the telephone line is 
working.
Transmission stopped 
unexpectedly. X of N ECGs sent. 
Remote site stopped 
communication (nn). Press any 
key to continue.
Communication speed of the 
remote device does not match 
that of the cardiograph, or 
remote site modem malfunction.
Call the remote location to 
verify that the communication 
speed is correct and that the 
modem is functioning correctly. 
Reduce the communication 
speed of both the cardiograph 
and remote site modems.
Table 7-5 Transmission Problems and Solutions(Continued) 
Message Likely Cause Possible Solutions