User`s guide
Filters
B-12 Configuring Your Cardiograph
Filters
Your PageWriter 200/300pi cardiograph has been factory-configured with the 
filter settings which remove the most noise from the ECG. In addition to the 
default settings, your cardiograph offers a choice of several filter configura-
tions. These filter settings are detailed below and in Table B-3.
l  The 0.5 Hz Baseline Wander filter suppresses the greatest amount of 
baseline wander. However, in Manual mode, this filter may alter the 
ECG’s ST segment. 
l The 0.15 Hz Baseline Wander filter provides some baseline wander 
suppression without distorting the ECG’s ST segment.
l The 0.05 Hz Baseline Wander filter delivers the highest fidelity sig-
nal, but provides the least baseline wander suppression. 
l The 40 Hz Noise filter offers maximum noise suppression, but 
reduces the fidelity of the signal.
l The 100 Hz Noise filter provides some noise suppression while offer-
ing an accurate signal representation. 
l The 150 Hz Noise filter delivers the highest fidelity signal, but pro-
vides the least high-frequency noise suppression. 
l The Artifact filter may be enabled for Auto and Manual ECGs when 
the   key is pressed. It removes small-amplitude, high-frequency 
signals, characteristic of muscle tremor.
Custom Rhythm Leads 
(Cabrera) for:
 3x4, 1R
 3x4, 3R, Manual 3
 Manual 6
I, II, III, -aVR, aVL, aVF, 
V1, V2, V3, V4, V5, V6
I, II, III, -aVR, aVL, aVF, 
V1, V2, V3, V4, V5, V6
I, II, III, -aVR, aVF, aVL, 
V1, V2, V3, V4, V5, V6
Select any lead
Select any three 
leads
Select any six 
leads
a. Default values are shown in boldface type.
Table B-2: Configurable Report Characteristics
Parameter
Choices
a
Comments
)LOWHU