NOTE When the   key is on, the user-configured combination of the Artifact, 
0.5 Hz Wander, and 40 Hz Noise filters is enabled and the Filter status 
message appears in the upper-right corner of the display. Refer to Table B-3 
for information on configuring the   key.
NOTE If accurate ST segment contours are required for ECGs recorded in Manual 
mode, do not use the 0.5 Hz baseline wander filter. This filter suppresses base-
line wander to the extent that it may alter the ST segment. Instead, configure 
your cardiograph to use the 0.15 Hz or 0.05 Hz baseline wander filter. Regard-
less of the filter used, the rhythm characteristics of the ECG are accurately 
recorded. 
)LOWHU
)LOWHU
Table B-3: Configurable Filters
Parameter
Choices
a
a. Default values are shown in boldface font.
Comments
Auto Baseline Wander Fil-
ter
0.05 Hz
0.15 Hz
0.5 Hz
Select one baseline 
wander filter for Auto 
ECGs.
Auto Noise Filter 40 Hz
100 Hz
150 Hz
Select one noise filter 
for Auto ECGs.
Manual Baseline Wander 
Filter
0.05 Hz
0.15 Hz
0.5 Hz
Select one baseline 
wander filter for 
Manual ECGs.
Manual Noise Filter 40 Hz
100 Hz
150 Hz
Select one noise filter 
for Manual ECGs.
 Artifact filter
 0.5 Hz Baseline Wander 
filter
 40 Hz Noise Filter
Yes  or No
Yes  or No
Yes  or No
Select whether or not 
the artifact, baseline 
wander and noise fil-
ters are activated by 
the   key.