User`s guide
Recording a Manual ECG
PageWriter 200/300pi M1771A/1770A Cardiograph 2-11
2
ID, Continue? is displayed. Press   to record the ECG 
without a patient ID. Press   to enter patient ID information.
l If you did not enter a patient ID in the above step but have entered 
patient ID information since the cardiograph was turned on, the 
question 
Continue using Old ID? is displayed. 
If you have entered a patient identification number and want to take 
more ECGs from the same patient, press  . 
If you are recording an ECG from another patient, press   
or .
NOTE Your cardiograph automatically checks for a patient ID each time you start an 
ECG, unless it is configured for no patient ID information. To bypass this 
check and use the last entered patient ID, press   a second time.
6. The cardiograph prints the ECG continuously until you press the   
key. 
NOTE If accurate ECG ST contours are required for ECGs recorded in Manual 
mode, do not use the 0.5 Hz baseline wander filter. This filter suppresses 
baseline wander to the extent that it may alter the ST segment. Instead, 
configure your cardiograph to use the 0.15 Hz or 0.05 Hz baseline wander 
filter. Regardless of the filter used, the rhythm characteristics of the ECG are 
accurately recorded. Refer to Configuring Your Cardiograph for details on 
configuring filters.
The leads recorded by the manual ECG, the chart speed, sensitivity, and filter 
status can be changed at any time, even while a manual ECG is running. To 
change these settings simply select  ,  ,  ,  , or   
keys, as appropriate. Note, however, that the ECG printer will pause as the 
cardiograph changes its recording status.
l The   key sequence is: 3 ⇒ 6 ⇒ 12 ⇒ 3 ... 
Ye s
No, Enter ID
Ye s
No, Edit Old ID
No, Enter New ID
0DQXDO
6WRS
Format Leads Speed Size
)LOWHU
Format










