Specifications
5-48
System Command Reference Section
System Level Status and Control
SYSTem:KLOCk <boolean>
This locks the instrument's keyboard. When locked, the user may not modify any of the
instrument’s configuration; although those keys that merely affect the display are still
usable.
The *RST selection is OFF.
SYSTem:PRESet|:PRESet<n>
Sets the error detector to a pre-defined “local operation” state. The choice of <n> is 0
through 2. PRESet and PRESet0 both have the same effect as the front-panel
key. PRESet1 and PRESet2 have the same effect as the front-panel
and keys respectively.
This command causes all past results to be labelled as invalid.
NOTE The GP-IB bus will be held off for approximately 8 seconds following
receipt of this command to allow the instrument to settle fully.
NOTE If PRESet2 is selected whilst an external controller is connected, then
an error message is given because the instrument is attempting to take
over GP-IB whilst it is already under control of the controller.
SYSTem:VERSion?
This command queries the version of the SCPI programming Language that the pattern
generator conforms to. The command currently returns 1990.0.
SYSTem:DATE <year>,<month>,<day>
This command permits the date of the real-time clock within the error detector to be set
up. The command is invalid if another MMS module, with an HP-MSIB address to the left
of and on the same row or lower than the error detector, exists and supports a valid date.
The range of valid <year> is 1990 through 2049.
SYSTem:TIME <hour>,<minute>,<second>
This command permits the time of the real-time clock within the error detector to be set
up. The command is invalid if another MMS module, with an HP-MSIB address to the left
of and on the same row or lower than the error detector, exists and supports a valid time.
INST PRESET
recall setup Preset 1 Preset 2