User`s guide

102 Chapter 4
Performance Tests
Performance Test Procedures
Performance Tests
Table 4-2 Recommended Test Equipment for 11974Q
Instrument Critical Specifications Recommended
Model
Spectrum Analyzer LO and IF ranges compatible with 11974Q
SWEEP + TUNE voltage compatible with
11974Q.
8566B
w/serial prefix 2948
or later
Synthesized Sweeper Frequency: 11 to 17 GHz
External leveling capability
Sub harmonics and multiples thereof: <
25 dBc
8340B
a
Amplifier Frequency Range: 3 to 5 GHz
Output level: +16 dBm leveled
11975A
Amplifier 15 dB gain from 11 to 17 GHz
Compatible with mm-wave source module
Non-harmonic spurious: <
50 dBc
8349B
Power Meter Compatible with thermistor mount 432A
Thermistor Mount SWR: <2.0 from 33 to 50 GHz Hughes
45772H-1100
Power Sensor SWR: <1.2 at 5 GHz 8481A
Power Meter Compatible with power sensor 436A
Directional Coupler Coupling: 10 dB
Directivity: >40 dB
Primary line SWR: <1.05
Auxiliary arm SWR: <1.1
Q752C
Isolator Insertion Loss: <1.6 dB
Isolation: >25 dB
SWR: <1.4
Q365A
Millimeter-Wave
Source Module
Output power: +2 dBm leveled
Output SWR: <2.0
Harmonically related spurious signals: <
20 dBc
83555A
Waveguide Straights 5 cm minimum
SWR: <1.016
Q896B
a. If an 8340A is used, the 1.0 V/GHz output will have to be modified to a
0.5 V/GHz output for proper display accuracy, leveling flatness, and harmonic
suppression. Provided with your mm-wave source module is a set of modification
instructions. This modification procedure will enable you to properly convert
your 8340A to a 0.5 V/GHz output.