User`s guide

66 Chapter 3
Operation
Using an 71000 Series Spectrum Analyzer with an 70907B, 70909A, or 70910A
Operation
Operation
Band Selection
1. If necessary, mate the 11974 to the spectrum analyzer being used for preselected
external mixing by using the procedure Frequency Tracking Calibration on page 64.
2. Use the following key sequence to enter the external mixing mode and to select the
desired band of operation.
MENU
State
ext mixer
fulband KAQUVE
Press the softkey for the desired band.
NOTE The above band selection key sequence must be used to enter the external
mixing band of operation. Do not enter in the center frequency of the band
directly.
Amplitude Calibration for a Single Frequency
Enter the conversion loss at the desired frequency from either the calibration label on the
bottom of the 11974 or the supplied calibration sheet by using the following key sequence.
State
ext mixer
CONV LOSS
Enter the conversion loss value of the desired frequency, then terminate the entry with
dB.
NOTE The above procedure is a single-point conversion-loss correction. For an
alternative procedure, available by remote programming only, refer to
Conversion Loss Versus Frequency Correction on page 69.