Specifications

Table 8-5. Status Messages (10 of 11)
The system was trying to access the following file when the error occurred: This is the
standard fiier error message.
Action:
Correct the problem and then continue.
Timed-out setting () on the... The System attempted to address an instrument, but the
instrument did not respond.
Action:
Access the System Configuration Table and and confirm that the correct HP-!B
addresses have been entered and that the instrument is responding.
Timed-out reading the ()... The System attempted to address an instrument but the instrument
did not respond.
Action:
Access the System Configuration Table and confirm that the correct HP-IB addresses
have been entered and that the instrument is responding.
Unable to Automatically Zero Beat the sources. Establish a beatnote less than or
equal to () Hz... The System was unable to adjust the beatnote to within 5% of the Peak
Tuning Range.
Action:
Adjust the source frequency until the beatnote is within 5% of the PTR.
Unable to integrate over non-contiguous data... Segments of the data are missing within the
specified Integration Range.
Action:
Verify the Segment Table ranges and remeasure the device across the entire desired
integration range.
Updating file directory... The System is modifying the directory within the mass storage media.
Use KNOB or ARROW KEYS to locate desired HP11848A internal path.
Action:
The user selects these keyboard controls to make selections from a menu.
Use KNOB or ARROW KEYS to locate desired test.
Action:
The user selects these keyboard controls to make selections from a menu.
VCO Tuning Constant: This message momentarily displays the measured VCO Tuning
Constant.
Verifying phase lock... The System is checking for the present of a beatnote on an
out-of-lock condition.
8-38 Status Messages