Oscilloscopes User's Guide

198 6000 Series Oscilloscope User’s Guide
4 Triggering the Oscilloscope
Table 1 4 Half-field holdoff time
Figure 20 Using TV Holdoff to Synchronize to Color Field 1 or 3 (Field 1
mode)
Standard Time
NTSC 8.35 ms
PAL 10 ms
PAL-M 10 ms
SECAM 10 ms
Generic 8.35 ms
EDTV 480p/60 8.35 ms
HDTV 720p/60 8.35 ms
HDTV 1080p/24 20.835 ms
HDTV 1080p/25 20 ms
HDTV 1080i/50 10 ms
HDTV 1080i/60 8.35 ms