Measurement Category
Measurement Category
The InfiniiVision 7000 Series oscilloscope is intended to be used 
for measurements in Measurement Category I.
Measurement Category Definitions
Measurement category I is for measurements performed on 
circuits not directly connected to MAINS. Examples are 
measurements on circuits not derived from MAINS, and 
specially protected (internal) MAINS derived circuits. In the 
latter case, transient stresses are variable; for that reason, the 
transient withstand capability of the equipment is made known 
to the user.
Measurement category II is for measurements performed on 
circuits directly connected to the low voltage installation. 
Examples are measurements on household appliances, portable 
tools and similar equipment.
Measurement category III is for measurements performed in the 
building installation. Examples are measurements on 
distribution boards, circuit-breakers, wiring, including cables, 
bus-bars, junction boxes, switches, socket-outlets in the fixed 
installation, and equipment for industrial use and some other 
equipment, for example, stationary motors with permanent 
connection to the fixed installation.
Measurement category IV is for measurements performed at the 
source of the low-voltage installation. Examples are electricity 
meters and measurements on primary overcurrent protection 
devices and ripple control units.
WARNING
Use this instrument only for measurements within its specified 
measurement category.