Technical data
22 7000 Series Oscilloscopes Service Guide 
2 Testing Performance
Overview
To completely test and troubleshoot MSO models, create and 
use the test connector accessory as described in this chapter.
• The test connector is only required for oscilloscopes that 
have the MSO option licensed (enabled).
• The connector is used in the digital channel threshold 
accuracy test.
• The test connector keeps electrical distortion to a minimum 
and makes it easy for you to connect the oscilloscope probes 
to function generators and measurement equipment. 
Let the Equipment Warm Up Before Testing
For accurate test results, let the test equipment and the 
oscilloscope warm up 30 minutes before testing.
Verifying Test Results
During the tests, record the readings in the Performance Test 
Record on page 57. To verify whether a test passes, verify that 
the reading is within the limits in the Performance Test Record.
If a performance test fails
If a performance test fails, first perform the User Cal procedure. Press the 
following keys to access User Cal: Utility→Service→Start User Cal.
7000 service guide.book Page 22 Thursday, October 16, 2008 12:18 PM










