Technical data
26 7000 Series Oscilloscopes Service Guide 
2 Testing Performance
To construct the test connector (for use with MSO models only)
Agilent 7000 Series Oscilloscopes that have digital channels 
enabled require the test connector described below. Follow the 
steps to build the test connector.
1 Obtain a BNC connector and an 8-by-2 section of Berg strip. 
A longer strip can be cut to length using wire cutters.
2 On one side of the Berg strip, solder a jumper wire to all of 
the pins (shown in Figure 1 on page 27).
3 On the other side of the Berg strip, solder another jumper 
wire to all of the pins.
4 Solder the center of the BNC connector to a center pin on one 
of the rows on the Berg strip.
5 Solder the ground tab of the BNC connector to a center pin 
on the other row on the Berg strip.
Table  3 Materials required to construct the test connectors
Description Recommended Part Qty 
BNC (f) Connector Agilent 1250-1032 or 
Pomona 4578
1
Berg Strip, 8-by-2 3M .100” x .100” Pin 
Strip Header or similar
1 strip, cut to length (8x2)
Jumper wire
7000 service guide.book Page 26 Thursday, October 16, 2008 12:18 PM