Technical data
Testing Performance 2
7000 Series Oscilloscopes Service Guide  29
To verify digital channel threshold accuracy (MSO models only)
This test verifies the digital channel threshold accuracy 
specification of the Agilent 7000 Series Oscilloscopes.
Threshold accuracy test limits = ±(100 mV + 3% of threshold 
setting)
When to Test
You should perform this test every 12 months or after 2000 
hours of operation, whichever comes first.
What to Test
Use these instructions to test the threshold settings of digital 
channels D7-D0. Then, use the same instructions to test digital 
channels D15-D8. 
Verifying Test Results
After each threshold test, record the voltage reading in the 
Performance Test Record on page 57. To verify whether a test 
passes, verify that the voltage reading is within the limits in the 
Performance Test Record.