Technical data
30 7000 Series Oscilloscopes Service Guide 
2 Testing Performance
1 Turn on the test equipment and the oscilloscope. Let them 
warm up for 30 minutes before starting the test. 
2 Set up the oscilloscope calibrator. 
a Set the oscilloscope calibrator to provide a DC offset 
voltage at the Channel 1 output. 
b Use the multimeter to monitor the oscilloscope calibrator 
DC output voltage. 
3 Use the 8-by-2 test connector and the BNC cable assembly to 
connect digital channels D0-D7 to one side of the BNC Tee. 
Then connect the D0-D7 ground lead to the ground side of 
the 8-by-2 connector. See Figure 2.
Table  4 Equipment Required to Test Digital Channel Threshold 
Accuracy
Equipment Critical 
Specifications
Recommended Model/Part 
Digital Multimeter 0.1 mV resolution, 
0.005% accuracy
Agilent 34401A 
Oscilloscope Calibrator DC offset voltage 
6.3 V
Fluke 5820A
BNC-Banana Cable Agilent 11001-66001 or 
Pomona 2BC-BNC-36
BNC Tee Agilent 1250-0781 or 
Pomona 3285
50
 Ω BNC Cable Agilent 10503A
BNC Test Connector, 
8-by-2
User-built (See page 26)
Probe Cable Agilent 01650-61607 or
Agilent 54620-68701
7000 service guide.book Page 30 Thursday, October 16, 2008 12:18 PM