Technical data
34 7000 Series Oscilloscopes Service Guide 
2 Testing Performance
1 Set up the oscilloscope. 
a Adjust the channel 1 position knob to place the baseline at 
0.5 major division from the bottom of the display. 
b Set the Volts/Div setting to the value in the first line in 
Table 7. 
Table  6 Equipment Required to Verify Voltage Measurement Accuracy
Equipment Critical Specifications Recommended 
Model/Part 
Oscilloscope 
Calibrator
14 mV to 35 Vdc, 
0.1 V resolution
Fluke 5820A
Digital multimeter Better than 0.01% accuracy Agilent 34401A 
Cable BNC, Qty 2 Agilent 10503A 
Shorting cap BNC Agilent 1250-0774 
Adapter BNC (f) to banana (m) Agilent 1251-2277 
Adapter  BNC tee (m) (f) (f) Agilent 1250-0781 or 
Pomona 3285
Blocking capacitor Agilent 10240-70001 + 
Agilent 1250-0774 or 
Agilent 11742A + Pomona 
4288 + Pomona 5088A