Technical data
Testing Performance 2
7000 Series Oscilloscopes Service Guide  35
c Press the Acquire key. Then press the Averaging softkey and 
set #Avgs to 64.
Wait a few seconds for the measurement to settle.
2 Press the Cursors key, set the Mode softkey to Normal, then 
press the XY softkey and select Y. Press the Y1 softkey, then 
use the Entry knob (labeled   on the front panel) to set 
the Y1 cursor on the baseline of the signal. 
3 Use the BNC tee and cables to connect the oscilloscope 
calibrator /power supply to both the oscilloscope and the 
multimeter (see Figure 3). 
Table  7 Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting  Oscilloscope 
Calibrator Setting
Tes t Limi ts
5 V/Div 35 V 34.04 V to 35.96 V 
2 V/Div 14 V 13.616 V to 14.384 V 
1 V/Div 7 V 6.808 V to 7.192 V 
500 mV/Div 3.5 V 3.404 V to 3.596 V 
200 mV/Div 1.4 V 1.3616 V to 1.4384 V 
100 mV/Div 700 mV 680.8 mV to 719.2 mV 
50 mV/Div 350 mV 340.4 mV to 359.6 mV 
20 mV/Div 140 mV 136.16 mV to 143.84 mV 
10 mV/Div 70 mV 68.08 mV to 71.92 mV 
5 mV/Div 35 mV 34.04 mV to 35.96 mV 
2 mV/Div
1
14 mV 13.232 mV to 14.768 mV 
1
 Full scale is defined as 32 mV on the 2 mV/div range. 
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
7000 service guide.book Page 35 Thursday, October 16, 2008 12:18 PM










