Technical data
56 7000 Series Oscilloscopes Service Guide 
2 Testing Performance
2 Set up the oscilloscope.
a Press the Save/Recall key, then press the Default Setup 
softkey.
3 Change the signal generator output frequency to 500 MHz.
4 Set the power meter Cal Factor % to the appropriate value 
(500 MHz) on the calibration chart on the power sensor. If 
necessary, do a linear interpolation if a 500 MHz factor is not 
included in the power meter’s calibration chart.
5 Adjust the signal generator output for reading on the power 
meter of 625µW. (500mV
pp
 = 176.78mV rms, 
Power = Vin
2
/50Ω = 176.78 mV
2
/50Ω = 625µW.)
6 Press the Trigger Edge key, then press the Source softkey to 
set the trigger source to External.
7 Check for stable triggering and adjust the trigger level if 
necessary. Triggering is indicated by the Trig’d indicator at 
the top of the display. When it is flashing, the oscilloscope is 
not triggered. When it is not flashing, the oscilloscope is 
triggered.
8 Record the results as Pass or Fail in the Performance Test 
Record (see page 57).
If the test fails, see the “Troubleshooting” chapter. Then 
return here.