Technical data
82 7000 Series Oscilloscopes Service Guide 
4 Troubleshooting
• Trigger - Edge trigger, Auto sweep mode, 0 V level, channel 
1 source, dc coupling, rising edge slope, 60 ns holdoff time
• Display - Vectors on, 33% grid intensity, infinite persistence 
off
• Other - Acquire mode normal, Run/Stop to Run, cursor 
measurements off
Figure 15 Default setup screen
3 If your screen looks substantially different, replace the 
system board.
7000 service guide.book Page 82 Thursday, October 16, 2008 12:18 PM










