User`s guide
4 Triggering
102 InfiniiVision Oscilloscopes User’s Guide
Duration Trigger
Duration trigger lets you define a pattern, then trigger on a specified time duration of this 
logical AND combination of the channels.
1 Press the [More] key in the Trigger section of the front panel, rotate the Entry knob 
until Duration is displayed in the Trigger softkey, then press the Settings softkey to 
display the Duration Trigger Menu.
2 For each analog or digital channel you want to include in the desired pattern, press the 
Channel softkey to select the channel.
This is the channel source for the H, L, or X condition. As you press the Channel 
softkey (or rotate the Entry knob), the channel you select is highlighted in the 
Pattern = line directly above the softkeys and in the upper-right corner of the 
display next to “Dur”. External trigger may also be specified as a channel in the 
pattern when using the 2-channel and 4-channel oscilloscopes.
Adjust the trigger level for the selected analog channel by turning the Trigger Level 
knob. Press the [D15-D0] key and select Thresholds to set the threshold level for 
digital channels. The value of the trigger level or digital threshold is displayed in the 
upper-right corner of the display.
3 For each channel you select, press the logic level softkey to set the condition for that 
channel in the pattern.
• 1 sets the pattern to 1 (high) on the selected channel. A high is a voltage level that 
is greater than the channel's trigger level or threshold level.
• 0 sets the pattern to 0 (low) on the selected channel. A low is a voltage level that is 
less than the channel's trigger level or threshold level.
Trigger level 
or threshold
Selected 
channel
Duration 
trigger
Channel 
select
Logic level
Return to 
previous menu
Qualifier
Qualifier time set