User`s guide
8 Acquisition Modes
272 InfiniiVision Oscilloscopes User’s Guide
If the signal you are measuring has more than about 1 s of “dead time” (inactivity), 
consider selecting Normal trigger mode to prevent AutoTriggering. Press the 
[Mode/Coupling] key, then press the Mode softkey and select Normal.
Navigating segments
Press the Current Seg softkey and turn the Entry knob to display to the desired segment 
along with a time tag indicating the time from the first trigger event.
You can also press the Knob softkey to toggle between Current Segment and Time. When 
Current Segment is selected, the Horizontal position control knob moves the display to 
the desired segment. When Time is selected, the Horizontal position control knob moves 
the display horizontally just as it does when Segmented Memory mode is not selected. 
The Knob softkey also appears in the Horizontal Menu so you can make the same 
adjustments there.
Sample 
rate
Progress indicator