User`s guide
10 Serial Decode/Lister
300 InfiniiVision Oscilloscopes User’s Guide
Counters
• The FRAMES counter gives the total number of completed remote, data, overload, and 
active error frames. 
• The OVLD counter gives the total number of completed overload frames and their 
percentage of the total number of frames.
• The ERR counter gives the total number of completed active error frames and their 
percentage of the total number of frames. 
• The UTIL (bus load) indicator measures the percentage of time the bus is active. The 
calculation is done on 330 ms periods, approximately every 400 ms.
Example: If a data frame contains an active error flag, both the FRAMES counter and the 
ERR counter will be incremented. If a data frame contains an error that is not an active 
error it is considered a partial frame and no counters are incremented.
Overload 
Frame Count 
and Percent
Error Frame 
Count and 
Percent
Bus Utilization 
Percentage
Total Frame 
Count