Specifications
17
Characteristics of the 1146B 
current probe
Bandwidth* dc to 100 kHz (–3 dB)
Current 
range*
100 mV/A:100 mA to 
10 A peak 
10 mV/A:1 to 100 A peak
Output signal 1000 mV peak max
AC current 
accuracy* 
 Range
 Accuracy 
 Range
 Accuracy 
 Range
 Accuracy
100 mV/A 
(50 mA to 10 A peak) 
3% of reading ±50 mA 
10 mV/A 
(500 mA to 40 A peak) 
4% of reading ±50 mA 
10 mV/A 
(40 A to 100 A peak) 
15% max at 100 A 
Noise Range 10 mV/A: 480 V 
Range 100 mV/A: 3 mV
Insertion 
impedance 
0.01  (50/60 Hz)
Maximum 
working 
voltage
600 Vrms CAT II or 
300 Vrms CAT III
Maximum 
common 
mode voltage
600 Vrms CAT II or 
300 Vrms CAT III
Influence 
of adjacent 
conductor
< 0.2 mA/A AC 
Influence of 
conductor 
position
0.5% of reading at 1 kHz 
in jaw
Battery 9 V alkaline (NEDA 
1604A, IEC 6LR61)
Low battery Green LED on when 
≤ 6.5 V 
Battery life 55 hours typical
Note: Reference conditions 23 ±5 °C, 
(73.4 ±41 °C) 20 to 75% relative humidity, 
dc to 1 kHz, probe zeroed, 1-minute 
warmup, batteries at 9 V + 0.1 V, 
external magnetic field <40 A/m, no dc 
component, no external current carrying 
conductor, 1 M/100 pF load, conductor 
centered in jaw.
Characteristics of the 1147B/N2893A  
current probe
Bandwidth 
(–3 dB)
dc to 50 MHz (1147B) 
dc to 100 MHz (N2893A)
Risetime 7 ns or less
Maximum 
current 
(continuous)
15 A peak, 15 A DC, 
10 Arms (when two 
probes are used with 
InfiniiVision scope)
30A peak, 30 A DC, 
24 Arms (when one 
probe is used with 
InfiniiVision scope)
Maximum 
peak current 
(non-
continuous)
30 A peak (when two 
probes are used with 
InfiniiVision scope)
32 A peak (when one 
probe is used with 
InfiniiVision scope)
Output 
voltage rate
0.1 V/A
Amplitude 
accuracy
±1% rdg, ±10 mA (dc 
and 45 to 66 Hz, rated 
current)
Noise Equivalent to 2.5 mArms 
or less (for 20 MHz 
bandwidth measuring 
instrument)
Temperature 
coefficient 
for sensitivity
±2% or less (within a 
range of 0 to 40 °C or 
32 to 104 °F) 
Effect of 
external 
magnetic 
fields
Equivalent to a 
maximum of 20 mA (in 
a dc to 60 Hz, 400 A/m 
magnetic field)
Maximum 
rated power
3 VA (with rated 
current)
Maximum 
input voltage
300 V CAT I
Diameter of 
measurable 
conductors
5 mm dia. (0.2 in dia.)
Probe 
interface
AutoProbe interface
Cable 
lengths
Sensor cable: 
Appox. 1.5 m (59.0n in) 
Power supply cable: 
Appox. 1 m (39.4 in)
Maximum 
number 
of probes 
supported
2
Note: The above specifications are 
guaranteed at 23 ±3 °C (or 73 ±5 °F)
Clamp-on Current Probes (continued)
Characteristics of N2780B Series 
current probes
Bandwidth 
(–3 dB)
dc to 2 MHz (N2780B) 
dc to 10 MHz (N2781B) 
dc to 50 MHz (N2782B) 
dc to 80 MHz (N2783L) 
dc to 100 MHz (N2783A)
Maximum 
current 
(continuous)
500 A (N2780B) 
150 A (N2781B) 
30 A (N2782B/83B/83L)
Maximum 
peak current 
(non-
continuous)
700 A peak (N2780B) 
300 A peak (N2781B) 
50 A peak 
 (N2782B/83B/83L)
Maximum 
input voltage
300 V CAT I 
 (N2782B/83B/83L) 
300 V CAT III, 600 V 
CAT II (N2780B/81B)
Output 
voltage rate
0.01 V/A 
 (N2780B/81B) 
0.1 V/A 
 (N2782B/83B/83L)
Amplitude 
accuracy
±1.0 % rdg ±500 mA 
 (N2780B) 
±1.0 % rdg ±100 mA 
 (N2781B) 
±1.0 % rdg ±10 mA 
 (N2782B) 
±1.0 % rdg ±10 mA 
 (N2783B/83L)
Ordering information for Agilent 
current probes
1146B 100-kHz current probe
1147B 50-MHz current probe with 
AutoProbe interface
N2893A 100-MHz current probe with 
AutoProbe interface
N2780B 2-MHz current probe
N2781B 10-MHz current probe
N2782B 50-MHz current probe
N2783L 80-MHz current probe with 
5 m long cable
N2783B 100-MHz current probe
N2779A 3-channel power supply for 
N2780B/81B/82B/83B/83L
For more information about the 
N2780B Series current probes, refer 
to the Agilent N2780B Series current 
probe data sheet, literature number 
5989-6432EN. 
*  Characteristics marked with asterisks are 
specified performance. Others are typical 
characteristics.