Specifications
7
InfiniiMax Active Probes and Accessories
1130A-34A InfiniiMax high-performance active probe system
•  1.5 - 7 GHz InfiniiMax probe system
•  InfiniiMax probe amplifier supports 
both differential- and single-ended 
measurements for a more cost-
effective solution
•  Unrivaled InfiniiMax probing 
accessories support browsing, 
solder-in, and socketed use models 
at the maximum performance 
available
•  Compatible with 3000 X-, 
4000 X-, 6000 X-, 5000, 6000, and 
7000 Series oscilloscopes (except 
for 6000 Series 100-MHz models)
The InfiniiMax 1130A Series probe 
amplifier is a perfect complement 
to the InfiniiVision 1 GHz or higher 
bandwidth models. Its bandwidth, 
extremely low input capacitance 
(0.32 pF), high common mode 
rejection, and patented resistor 
probe tip technology provide ultra 
low loading of the DUT and superior 
signal fidelity. Agilent’s innovative 
InfiniiMax 1130A Series differential 
probe is the easiest-to-use, and 
highest performance probing system 
available for high-speed digital 
design, and represents a new industry 
standard for accuracy, flexibility, and 
reliability. Designers can achieve full 
system bandwidth in conjunction 
with InfiniiVision Series oscilloscopes 
even when manually browsing 
with the probe or making hands-off 
measurements. Optional solder-in 
probe heads and solder-in sockets, as 
well as browser configuration provide 
full bandwidth at the probe tip.
Operating characteristics
Probe bandwidth (–3 dB) 1134A: > 7 GHz
1132A: > 5 GHz
1131A: > 3.5 GHz
1130A: > 1.5 GHz
Rise and fall time 
(10 to 90%)
1134A: 60 psec
1132A: 86 psec
1131A: 100 psec
1130A: 233 psec
Input capacitance Cm = 0.1 pF Cm is between tips 
Cg = 0.34 pF Cg is ground for each tip 
Cdiff = 0.27 pF Differential mode capacitance = Cm + Cg/2 
Cse = 0.44 pF Single-ended mode capacitance = Cm + Cg
Input resistance Differential mode resistance = 50 k ± 2% 
Single-ended mode resistance = 25 k ± 2%
Input dynamic range ±2.5 V
Input common mode range ±6.75 Vdc to 100 Hz; ±1.25 V > 100 Hz
Maximum signal slew rate 18 V/ns when probing a single-ended signal 
30 V/ns when probing a differential signal
DC attenuation 10:1 ± 3% before calibration on oscilloscope 
10:1 ± 1% after calibration on oscilloscope
Offset range ±12.0 V when probing single-ended signal
Maximum input voltage 30 Vpeak, CAT I
ESD tolerance > 8 kV from 100 pF, 300  HBM
Maximum number of 
probes supported by 
3000 X-/5000/6000/7000 
Series
2
Maximum number of 
probes supported by
4000 X- and 6000 X-Series
4
Agilent 1130A InfiniiMax probe offers you the highest performance available for 
measuring differential and single-ended signals