Technical data
N2780/81/82/83 User’s and Service Guide 17
This demagnetizes the core if it has been magnetized by 
switching the power on and off, or by an excessive input. 
Always carry out demagnetizing before measurement. The 
demagnetizing process takes about one second. During 
demagnetizing, a demagnetizing waveform is output. 
4 Zero adjustment dial (ZERO ADJ) 
Use the zero adjustment dial to correct for the effect of a 
voltage offset or temperature drift on the unit. The probe 
should be always be zeroed after demagnetization. 
5 Coarse adjustment trimmer 
This adjustment should only be carried out if the probe 
offset is outside the range of the zero adjustment dial. 
6 BNC output connector 
Connect to the BNC input connector of the oscilloscope. 
7 Power plug 
Connect this to the N2779A power supply receptacle to 
supply power to the sensor terminator. 
8 Barrier
This structure reduces the likelihood of touching the 
conductor while testing, and indicates the limit of safe 
physical contact. Avoid touching the clamp in front of the 
barrier when clamping or measuring.
NOTE
The output of the current probe is terminated internally. 
You must select the input impedance of the oscilloscope to 
be 1 M in order to make accurate measurements. If the 
oscilloscope you are using does not have a 1 M input 
impedance setting you can purchase the Agilent E2697A 
50 to 1 M adapter.
Turn the collar until it clicks, and check that it is locked 
securely.
medium_standard.book Page 17 Friday, September 10, 2010 2:45 PM










