Technical data
36 N2780/81/82/83 User’s and Service Guide
NOTE
• The output of this unit is terminated internally. Use an 
oscilloscope with an input impedance of at least 1 M or 
use the Agilent E2697A impedance adapter. 
• Immediately after powering on the probe, the probe may 
be subject to an appreciable offset drift due to the effect 
of self- heating. To counteract this, allow the probe to 
warm up for about 30 minutes before carrying out 
measurement. 
• When performing continuous measurements, it is 
necessary to be aware that the offset voltage drifts, 
depending on factors such as the ambient temperature. 
• Under certain circumstances, oscillation may occur if the 
probe is connected to the N2779A power supply while the 
power supply is on. This does not indicate a malfunction. 
Oscillation can be stopped and operation restored to 
normal by opening and closing the sensor head. 
• Depending on the measured current frequency, some 
sound maybe produced by resonance, but has no effect on 
measurements. 
• The reading may be affected by the position within the 
clamp aperture of the conductor being measured. The 
conductor should be in the center of the clamp aperture. 
• When carrying out a measurement, press the opening 
lever until the “UNLOCK” indication disappears and check 
that the sensor head is properly closed. If the sensor head 
is not properly closed, an accurate measurement is not 
possible. 
• Accurate measurement may be impossible in locations 
subject to strong external magnetic fields, such as 
transformers and high- current conductors, or in locations 
subject to strong external electric fields, such as radio 
transmission equipment. .