PRBS Generator 10 Gb/s User Guide 
Figure 7.  Verification setup 
Correct pattern length is determined by confirming that the tone spacing on 
the spectrum analyzer equals (Output_Bit_Rate/(2
n
-1)), where n is the 
specific pattern length. As an example, at 10 Gb/s the tone spacing for a  
(2
31
-1) pattern should be ~ 4.66 Hz (refer to Table 4 for tone spacings).  