26 User’s Guide U3025-90001
U3025AE10 Controlling the System with E8363/4C and N5230C
1. On the network analyzer select Response > Cal Wizard. 
a. If using a mechanical cal kit, select SmartCal (Guided Calibration) > Next.
b. If using an ECal module, connect the ECal to an available PNA USB port and select 
Use Electronic Calibration (ECal) > Next.
2. Continue following the Cal Wizard prompts. In the Select Calibration Ports and ECal 
Module dialog box press Select All, or select the ports you are calibrating and select 
Next. 
Figure 26 12-Port Calibration
3. Connect the ECal or the mechanical cal kit to the ports you are calibrating following the 
Cal Wizard prompts and press Measure after each connection. 
4. At the Calibration Completed prompt select Save As User Calset, type the name and 
Save. See Figure 32 on page 34.
5. After calibrating Test Set ports, use a quality load and short to verify the calibration on 
each port or end of the test cable. Measure reflection and confirm the return loss is as 
expected. If the result is not as expected, repeat the calibration without the Test Set and 
ensure that the PNA is in standard (non-multiport) mode. 
NOTE If measurement errors occur, ensure the newest version of firmware is 
installed on the PNA. Measurement errors can be a result of firmware 
algorithms. Consult with Agilent Service or firmware web page for the latest 
PNA Option 551 firmware revisions and history. 
http:\\na.tm.agilent.compna\firmware\firmware.htm.