User’s Guide U3025-90001 37
 U3025AE10 Operational Check Procedures
10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in Table 9 on 
page 32. This can be done using Limit Lines, select Marker/Analysis > Analysis > Limit 
Test > Limit Test ON and Limit Line ON > Global pass/fail display ON, select Show 
Table . The trace should be above the limit. PASS will be displayed on the screen if the 
limit lines are used. 
11.Repeat step 9 and step 10 for Cal Sets “999.1” thru “999.12” (12-Port).
NOTE For Ports 3-12 the response from 10 MHz to 500 MHz is normal due to the 
PNA-X Couplers in comparison to the U3022AE10 bridges. The bridges have 
more gain in the coupled RF path. 