Technical data
3-26 Service Guide N5230-90025
Tests and Adjustments PNA Series Microwave Network Analyzers
System Verification N5230C
Interpreting the Verification Results
Figure 3-11 shows an example of typical verification results with Print Tabular Data selected in the Printer 
Output area of the System Verification dialog box.
At the top of the printed output is the name of the device, the serial number of the device, and the date 
tested.
Each S-parameter measurement result is printed with frequency tested, lower and upper limit lines, the 
measured data, and the result of the test.
Figure 3-11 Example of Printed Tabular Verification Results
Sys Ver -- 20 dB attenuator magnitude results, Serial #00810 -- 6 Sep 2000
S11 Results
Frequency  Lower Limit  Measured Data Upper Limit Result
 0.3 MHz  0.0048 Units  0.0090 Units 0.0139 Units PASS
 100  0.0046  0.0091  0.0137  PASS
 200  0.0042  0.0092  0.0134  PASS
 300  0.0040  0.0091  0.0133  PASS
 400  0.0038  0.0089  0.0131  PASS
 500  0.0036  0.0087  0.0129  PASS
 600  0.0034  0.0085  0.0127  PASS
 700  0.0031  0.0082  0.0125  PASS
 800  0.0029  0.0080  0.0122  PASS
 900  0.0026  0.0079  0.0119  PASS
 1000  0.0023   0.0075  0.0117  PASS
 1100  0.0020   0.0072  0.0114  PASS
 1200  0.0017   0.0068  0.0111  PASS
 1300  0.0013   0.0064  0.0107  PASS
 1400  -0.0006   0.0059  0.0118  PASS
 1500  -0.0011   0.0052  0.0113  PASS
 1600  -0.0015   0.0044  0.0108  PASS
 1700  -0.0020   0.0038  0.0103  PASS
 1800  -0.0026   0.0032  0.0097  PASS
 1900  -0.0031   0.0024  0.0091  PASS
 2000  -0.0036   0.0019  0.0085  PASS
 2100  -0.0045   0.0016  0.0082  PASS
 2200  -0.0050   0.0018  0.0077  PASS
 2300  -0.0054   0.0023  0.0073  PASS
 2400  -0.0052   0.0031  0.0075  PASS
 2500  -0.0048   0.0039  0.0080  PASS
 2600  -0.0040   0.0050  0.0087  PASS
 2700  -0.0032   0.0060  0.0094  PASS
 2800  -0.0024   0.0071  0.0103  PASS
 2900  -0.0014   0.0082  0.0113  PASS
 3000  -0.0004   0.0095  0.0124  PASS
Overall PASS/FAIL result for entire frequency range = PASS










