Technical data
146 Notes on USB Electrical Compliance Testing
5 Host Hi-Speed Electrical Tests
3 Click  OK to close the Test Instructions dialog.
Test Instructions, Part 2
The USB automated test application will prompt you to perform these 
steps:
1 Connect the [TEST PORT] of the E2649- 66402 Host Hi- Speed Signal 
Quality test fixture (E2645- 66508 if you are using the old fixture) into 
the port under test of the host controller, using the 4" USB cable. The 
host controller here refers to the Hi- speed Electrical Test Bed Computer 
that has the HS Electrical Test Tool on it.
2 Select  TEST_PACKET from the Port Control drop down menu.
3 Enter the port number of the port under test and click [EXECUTE]. 
This forces the port under test to continuously transmit test packets.
You should see the transmitted test packet on the oscilloscope as below.