Technical data
196 Notes on USB Electrical Compliance Testing
7 Low and Full Speed Tests
4 Connect the Channel 1 probe to Vbus on the SQiDD board with the 
probe's ground to GND on one of the 100 mA load boards. This is the 
port under test.
5 Connect the Channel 2 probe to the Vbus test point on the load board 
connected adjacent to the USB port to be measured. This provides the 
oscilloscope trigger.
6 Click  OK to close the Test Instructions dialog.
Viewing Test Results
1 When the Testing Complete dialog appears, click OK.
The Results tab shows the test results.