Technical data
Low and Full Speed Tests 7
Notes on USB Electrical Compliance Testing 217
Connecting the Equipment  The USB automated test application will prompt 
you to perform these connection steps:
1 For 54831B/D and 54832B/D oscilloscopes, use 1165A passive probes. 
For the 5485XA, 80000 and 90000A Series oscilloscpes, use E2697A 
high- impedance converter with 10:1 passive probes, or 1156A active 
probes.
2 Attach the passive probes to the oscilloscope's channel Channel 2, 
Channel 3 and Channel 1 inputs.
3 Attach the SQiDD board to two USB ports at the end of 5 self- powered 
hubs and a host system. Hub #1 has to be a hi- speed hub and hub #2 
has to be a full speed hub. The rest of the hubs can be either hi- speed 
or full speed hubs.
4 Attach a full speed device under test to the same section of the SQiDD 
board. If the section has a switch, it should be set to ON.
5 Attach another full speed device to the adjacent section of the SQiDD 
board. This is for triggering purposes.
6 Connect the oscilloscope Channel 2 probe to D-  probe point of the 
device under test portion. Connect the oscilloscope Channel 3 probe to 
the D+ probe point of the device under test portion. Connect the 
oscilloscope Channel 1 probe to the D+ probe point on the adjacent 
device section of the SQiDD board.
NOTE
Placing a full speed and/or a high-speed device downstream of a full speed hub forces both 
to operate in full speed mode.
NOTE
Use a full-speed hub if device under test has embedded hub function. Otherwise the scope 
will falsely trigger.