Technical data
36 Notes on USB Electrical Compliance Testing
3 Device Hi-Speed Tests
Configuring the Tests
Connecting the Equipment - Differential Connection
The USB automated test application will prompt you to perform these 
connection steps:
1 Attach the 5V power supply to J5 of the E2649- 66401 (E2645- 66507 if 
you are using the old fixture) Device Hi- Speed signal quality test 
fixture. Leave the TEST switch at the OFF position. Verify the green 
Power LED is lit and the yellow Test LED is not lit.
2 Connect the [TEST PORT] of the Device Hi- speed Signal Quality test 
fixture into the upstream facing port of the device under test, using the 
4" USB cable.
3 Connect the [INIT PORT] of the test fixture to a Hi- speed capable port 
of the Test Bed Computer, using a USB cable.
4 Apply power to the device.
5 Attach the differential probe to D+/D- of TP2 on the test fixture, using 
the damped header adapter (the header adapter is only needed if you 
are using the old test fixture). Ensure the + polarity on the probe lines 
up with D+.
6 If you are using the new fixture, please terminate the SMA connectors 
with 50 Ohm terminators.
7 Check  I have completed these instructions.