Technical data
Device Hi-Speed Tests 3
Notes on USB Electrical Compliance Testing 49
Configuring the Tests
Connecting the Equipment
The USB automated test application will prompt you to perform these 
connection steps:
1 Connect the E2697A with 10073C passive probe or the 1156A active 
probe on Channel 2 to the D-  pin at TP2 of the E2649- 66401 
(E2645- 66507 if you are using the old fixture) Device Hi- Speed Signal 
Quality test fixture.
2 Connect the E2697A with 10073C passive probe or the 1156A active 
probe on Channel 3 to the D+ pin at TP2.
3 Connect both probe grounds to GND leads.
4 Connect the [INIT PORT] of the test fixture into the HS host controller 
port, using the 5- meter USB cable.
5 Connect the [TEST PORT] on the fixture to the upstream port of the 
device under test. Do not apply 5V to the test fixture.
6 If you are using the new fixture, please terminate the SMA connectors 
with 50 Ohm terminators.
7 Check  I have completed these instructions.