Product specifications

29
Analog/resistance programming
For greater programming flexibility 
and to minimize the time associated 
with computer interaction, both the 
output voltage and current can be 
controlled by an external voltage 
or resistance programming signal.
Connect multiple units in 
parallel and series
Should you need greater output 
power, the N5700 series power  
supplies give you the flexibility 
to connect in parallel up to four 
similarly rated units for greater 
output current and connect two 
similarly rated units in series for 
greater output voltage (see DC 
float voltage information)
Fan speed control
Lab bench use is enhanced by the 
fan speed control, which helps to 
minimize the acoustic noise.
Application notes
Side-by-Side Comparison:  
Agilent N5700 System DC Source 
and Sorensen DLM Power Supply 
5989-1628EN
Side-by-Side Comparison: 
Agilent N5700 System DC Source 
and Xantrex XFR DC Power Supply
5989-1630EN
Trends in Programmable Medium Power
(~1 kW) System DC Power Supplies
5989-1331EN
  Single-Output 750 W & 1500 W GPIB, LAN, USB
(Continued)
Spe cif icati ons
(at 0° to 55°C unless 
otherwise specified)
More detailed specifications at 
www.agilent.com/find/N5700
Basic DC Power Supplies   essential features for a tight budget
    N5747A  N5748A  N5749A  N5750A  N5751A  N5752A 
Number of outputs    1  1  1  1  1          1
GPIB, LAN, USB    Yes  Yes  Yes  Yes  Yes  Yes
Ouput ratings
Voltage    60 V  80 V  100 V  150 V  300 V  600 V
Current    12.5 A  9.5 A  7.5 A  5 A  2.5 A  1.3 A
Power    750 W  760 W  750 W  750 W  750 W  780 W 
Programming accuracy
Voltage  0.05%+  30 mV  40 mV  50 mV  75 mV  150 mV  300 mV 
Current  0.1%+  12.5 mA  9.5 mA  7.5 mA  5 mA  2.5 mA  1.3 mA 
Output ripple and noise  
CV p-p
(Up to 20 MHz)    60 mV  80 mV  80 mV  100 mV  150 mV  300 mV
CV rms
(From 5 Hz – 1 MHz)    8 mV  8 mV  8 mV  12 mV  20 mV  60 mV
Readback accuracy 
Voltage  0.1%+  60 mV  80 mV  100 mV  150 mV  300 mV  600 mV 
Current 
0.1%+
  37.5 mA  28.5 mA  22.5 mA  15 mA  7.5 mA  3.9 mA 
Load regulation 
(change from 10% to 90%) 
Voltage    8 mV  10 mV  12 mV  17 mV  32 mV  62 mV
Current    7.5 mA  6.9 mA  6.5 mA  6 mA  5.5 mA  5.26 mA
Line regulation 
(change from 85-132 VAC input 
or 170-265 VAC input)   
Voltage    8 mV  10 mV  12 mV  17 mV  32 mV  62 mV
Current    3.25 mA  2.95 mA  2.75 mA  2.5 mA  2.25 mA  2.13 mA
Transient response time
1
Time    ≤ 1 ms  ≤ 1 ms  ≤ 1 ms  ≤ 2 ms  ≤ 2 ms  ≤ 2 ms
Supplemental Characteristics 
  Non-warranted characteristics determined by design  
                                              that are useful in applying the product)
Output response time 
(settle to within ±1.0% of the  
rated output, with a resistive load) 
Up, full load    0.08 s  0.15 s  0.15 s  0.15 s  0.15 s  0.25 s
Down, full load    0.08 s  0.15 s  0.15 s  0.15 s  0.15 s  0.30 s
Down, no load    1.1 s  1.2 s  1.5 s  2.0 s  3.0 s  4.0 s
Remote sense compensation
Volts/load lead     3 V  4 V  5 V  5 V  5 V  5 V
Output ripple and noise
2
CC rms     38 mA  29 mA  23 mA  18 mA  13 mA  8 mA
Programming resolution/ 
measurement resolution    
Voltage    7.2 mV  9.6 mV  12 mV  18 mV  36 mV  72 mV
Current    1.5 mA  1.14 mA  0.9 mA  0.6 mA  0.3 mA  0.156 mA
Notes: 
1
 Time for output voltage to recover within 
  0.5% of its rated output for a load change from 
  10 to 90% of its rated output current. Voltage 
  set point from 10% to 100% of rated output
2
 From 5 Hz – 1 MHz, at 10% to 100% of output 
  voltage at full load (for 6 V units from 33% to 
  100% of output voltage)