Product specifications

This reliable series of 500 W DC 
power supplies can be controlled 
either from the front panel or via 
an analog programming voltage. 
When used in a test system, the fast 
up and down programming helps 
decrease test time. Quickly reacting 
protection features, including fast 
crowbar, CV/CC mode crossover and 
over-voltage protection help protect 
your valuable assemblies from dam-
age. The linear topology produces 
very low ripple and noise, which 
allows you to make extremely accu-
rate measurements of the devices 
which you are testing.
Lab bench use is enhanced by the 
fan speed control, which helps to 
minimize the acoustic noise.
High Performance DC Power Supplies   speed and accuracy for test optimization
Spe cif icati ons
(at 0° to 55°C unless 
otherwise specified)
More detailed specifications at 
www.agilent.com/find/6550
52
Front panel and analog control of output voltage and current 
Fast, low-noise outputs
Fan-speed control to minimize acoustic noise
Protection features to ensure DUT safety
  Single-Output
  500 W   
    6551A  6552A   6553A  6554A   6555A   
Number of outputs    1  1  1  1  1 
GPIB    No  No  No  No  No 
Output ratings  
Output voltage    0 to 8 V   0 to 20 V  0 to 35 V   0 to 60 V   0 to 120 V
Output current (40°C)     0 to 50 A   0 to 25 A   0 to 15 A  0 to 9 A   0 to 4 A
Maximum current (50°C/55°C)    
45 A/42.5 A  22.5 A/21.3 A   13.5 A/12.8 A   8.1 A/7.7 A  3.6 A/3.4 A
Programming accuracy 
at 25°C ±5°C 
Voltage   0.06% +   5 mV   10 mV  15 mV   26 mV  51 mV
Current   0.15% +   60 mA  25 mA   13 mA   8 mA   4 mA
Ripple and noise 
from 20 Hz to 20 MHz 
Voltage rms     300 µV  300 µV  400 µV   500 µV   700 µV
peak-peak     3 mV  3 mV   4 mV  5 mV  7 mV
Current rms     25 mA  10 mA   5 mA   3 mA   2 mA
Load regulation 
Voltage     1 mV   2 mV  3 mV  4 mV  5 mV
Current     2 mA   1 mA  0.5 mA  0.5 mA  0.5 mA
Line regulation 
Voltage     0.5 mV  0.5 mV  1 mV  1 mV  2 mV
Current     2 mA   1 mA  0.75 mA  0.5 mA   0.5 mA
Transient response time    Less than 100 µs for the output voltage to recover to its previous level 
(within 0.1% of the voltage rating of the supply or 20 mV, whichever is greater)
    following any step change in load current of up to 50% of rated current  
 Supplemental Characteristics  (Non-warranted characteristics determined by design and 
    useful in applying the product)
Average resolution 
Voltage     2 mV  5 mV  10 mV  15 mV   30 mV
Current    15 mA  7 mA  4 mA  2.5 mA  1.25 mA
OVP     12 mV  30 mV  54 mV  93 mV   190 mV
OVP accuracy     160 mV  400 mV  700 mV  1.2 V   2.4 V
6551A-6555A