Stereo System User Manual
Chapter 14  175
1xEV-DO Measurement Application
Measurements
Description Specifications Supplemental Information
Spurious Emissions 
Dynamic Range, relative 91.9 dB 97.1 dB (typical)
Sensitivity, absolute −79.4 dBm −85.4 dBm (typical)
Accuracy, absolute
 20 Hz to 3.6 GHz  ±0.38 dB (95% confidence)
 3.5 GHz to 8.4 GHz  ±1.22 dB (95% confidence)
 8.3 GHz to 13.6 GHz ±1.59 dB (95% confidence)
Description Specifications Supplemental Information
QPSK EVM
−25 dBm ≤ ML
a
≤ −15 dBm 
 20 to 30 °C
a. ML (mixer level) is RF input power minus attenuation
RF input power range is accordingly 
determined to meet Mixer level.
EVM
 Operating range 0 to 25%
 Floor 1.5% 
 Accuracy
b
b. The accuracy specification applies when the EVM to be measured is well above the measurement floor. 
When the EVM does not greatly exceed the floor, the errors due to the floor add to the accuracy errors. 
The errors due to the floor are noise-like and add incoherently with the UUT EVM. The errors depend 
on the EVM of the UUT and the floor as follows: error = sqrt(EVMUUT
2
 + EVMsa
2
) – EVMUUT, 
where EVMUUT is the EVM of the UUT in percent, and EVMsa is the EVM floor of the analyzer in 
percent.
±1.0%
I/Q origin offset
 DUT Maximum Offset −10 dBc (nominal)
 Analyzer Noise Floor −50 dBc (nominal)
Frequency Error 
 Range
±30 kHz (nominal)
 Accuracy
±5 Hz + tfa
c
c. tfa = transmitter frequency x frequency reference accuracy.